Project SERVICE MANUAL
Date
05-2017
67/84
6.2.7.2 Positioning the weights for sensor 4 (right foot sensor)
6.2.8 Calculating and saving the new sensor constant - sensor 4:
sensor 4:
check the value on the secondary indicator
value of secondary display
x sensor 4 constant = new constant
value of weights used (80kg)
Write the new constant value
to the position of sensor 4
79,5
x 1.953 = 1,940
80