7 Errors – Failure States
7.1 Pop ups
Pop-Up
Meaning
Action required
Function locked
Activate function by unlocking respective
lock
X-Ray cassette holder not inserted correctly
Insert X-Ray cassette holder correctly
Activation required
Press GO button to activate keypad or
touchscreen
Siderails folded down – lateral tilt disabled
Fold up siderails to enable adjustment of
lateral tilt
Prevention of collision of bed and floor in
tilted position
Shorten mattress platform or lower
calfrest
Horizontal position reached during tilting of
bed
Press respective button or icon to
continue tilting
Lateral tilt adjustment by foot control – max.
15°
No action required - information only
Backrest angle adjusted above 30° - Lateral
tilt is limited to 15°
Lateral tilt is adjusted more than 15°-
backrest angle is limited to 30°
Lower backrest to an angle of less than
30° to adjust lateral tilt to an angle of
more than 15°
Adjust lateral tilt to an angle less than 15°
to raise the backrest to an angle of more
than 30°
Maximum load of exceeded
Remove weight
Bed loaded with more than 150 kg – Lateral
tilt only possible to 15°
No action required – information only
Function can not be activated when running
on battery.
Connect bed to mains
Patient transfer is activated (Lateral tilt only
possible to 15°)
No action required – Information only
Automatic lateral therapy is running.
Positioning of bed is disabled.
Stop automatic lateral therapy to enable
positioning of bed.
A potentiometer has malfunctioned
Service or replacement of column unit
neccesary
Function stopped due to possible collision
in lateral tilt.
Level the bed by continuing to press the
same button or manually (see …..)
Malfunction of bed movement protection
after perfoming one movement.
Service needed or replacement of
electrical parts of bed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LINET MULTICARE 2015 and is the answer not in the manual?

LINET MULTICARE 2015 Specifications

General IconGeneral
BrandLINET
ModelMULTICARE 2015
CategoryMedical Equipment
LanguageEnglish

Related product manuals