5.5.3 Activating / deactivating short-circuit detection on the ACP
To enter short-circuit detection switching mode:
Activate the ACP by pressing the GO button. Then
press mattress platform UP + DOWN + GO
simultaneously for 5s.
The LED mains power indicator will start blinking
and an acoustic signal will be heard.
Switching is achieved using the leg rest lock
button
Short-circuit detection is ON when the leg rest
lock is not lit.
Short-circuit detection is OFF when the leg rest
lock is lit.
Mark of ACP with short-circuit detection (ST)