Project SERVICE MANUAL
Date
05-2017
Version 05
55/84
5.7 Calibrating the i-Drive Power
®
accelerometers
It is necessary to calibrate the i-Drive Power® accelerometer after any service intervention. The bed must be on a
level floor. This can be checked using the spirit level on the head end of the bed.
Calibrate i-Drive Power
®
as follows:
Ensure the bed is on a level floor!
In case the i-Drive Power® system is not activated, activate it by pressing the activation button on the
undercarriage.
Activate the wheel at the head end of the bed so it lowers into the working position.
Press and hold the FWD and REV buttons simultaneously for 5s.
Dont touch „Safety sense“ sensor by hand
Hold the buttons FWD and REV and the system will enter the accelerometer calibration mode. The battery
status LED will be turned off.
The battery status LED will flash and a tone will be heard.
Release the FWD and REV buttons.
The i-Drive Power® accelerometer has been succesfully calibrated.