Channel Calibration
- 22 -
MXD70 Suspended Solids / Turbidity
Setup and Operating Guide
Channel Calibration
Defer Calibration Date
Turns off the alarm and increases the calibration
interval by an extra 7 days.
Only appears once the calibration interval has
expired.
YES
– Increase Interval
NO
– Cancel
Resetting the User Calibration
If required the user can reset the user calibrations to their default states.
Main Menu
From the front screen press the menu button to
show the main menu options and select Calibration.
/ – Select Option
EXIT
– Return to Front Screen
– Enter Option
Calibration
Select Reset User Calibration.
/ – Select Option
EXIT
– Return to Main Menu
– Enter Option
Reset User Calibration
Select the required Suspended Solids or Turbidity
input channel.
/ – Select Option
EXIT
– Return to Calibration
– Enter Option