18 - 4 Operator’s Manual
18 System Maintenance
Cleaning the dust-proof covers
Please clean all dust-proof covers of the system periodically (1 time per
month); otherwise, system damage may result. Cleaning times can be
increased when the system is used in the open air or somewhere dust is more.
Tool: soft brush
Method:
Disassemble dust-proof cover before cleaning.
18.1.2 Disinfecting the Main Unit
Only the system equipped with trackpad supports disinfection.
• Use only Mindray approved disinfectants and methods listed in this section
to disinfect the main unit. Warranty does not cover damage caused by
unapproved substances or methods.
• Do not mix disinfectants, as hazardous gases may result.
• We make no claims regarding the efficacy of the listed chemicals or
methods as a means for controlling infection. For the method to control
infection, consult your hospital’s infection control officer or epidemiologist.
• Never immerse any part of the main unit in liquids or allow liquid to enter
the interior.
• Any contact of disinfectants with connectors or metal parts may cause
corrosion.
• Do not pour or spray any liquid directly on the main unit or permit fluid to
seep into connections or openings.