© National Instruments | 6-17
X Series User Manual
When you generate digital output signals, the generation pauses as soon as the pause trigger is
asserted. If the source of your sample clock is the onboard clock, the generation resumes as soon
as the pause trigger is deasserted, as shown in Figure 6-9.
Figure 6-9. DO Pause Trigger with the Onboard Clock Source
If you are using any signal other than the onboard clock as the source of your sample clock, the
generation resumes as soon as the pause trigger is deasserted and another edge of the sample
clock is received, as shown in Figure 6-10.
Figure 6-10. DO Pause Trigger with Other Signal Source
Using a Digital Source
To use DO Pause Trigger, specify a source and a polarity. The source can be one of the following
signals:
• PFI <0..15>
• RTSI <0..7>
• PXI_STAR
• PXIe_DSTAR<A,B>
• Counter n Internal Output
• Counter n Gate
• AI Pause Trigger (ai/PauseTrigger)
• AO Pause Trigger (ao/PauseTrigger)
• DI Pause Trigger (di/PauseTrigger)
The source can also be one of several other internal signals on your DAQ device. Refer to Device
Routing in MAX in the NI-DAQmx Help or the LabVIEW Help for more information.
You can also specify whether the samples are paused when DO Pause Trigger is at a logic high
or low level.
Pause Trigger
Sample Clock
Pause Trigger
Sample Clock

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments 636 Series and is the answer not in the manual?

National Instruments 636 Series Specifications

General IconGeneral
BrandNational Instruments
Model636 Series
CategoryData Loggers
LanguageEnglish

Related product manuals