EasyManuals Logo
Home>National Instruments>Data Loggers>636 Series

National Instruments 636 Series User Manual

National Instruments 636 Series
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #62 background imageLoading...
Page #62 background image
© National Instruments | 4-19
X Series User Manual
When to Use Referenced Single-Ended (RSE) Connections with
Ground-Referenced Signal Sources
Do not use RSE connections with ground-referenced signal sources. Use NRSE or DIFF
connections instead.
As shown in the bottom-rightmost cell of Table 4-3, there can be a potential difference between
AI GND and the ground of the sensor. In RSE mode, this ground loop causes measurement
errors.
Using Differential Connections for Ground-Referenced Signal
Sources
Figure 4-10 shows how to connect a ground-referenced signal source to the MIO X Series device
configured in differential mode.
Figure 4-10. Differential Connections for Ground-Referenced Signal Sources
Note (NI USB-6341/6343/6346/6361/6363 BNC Devices) To measure a
ground-referenced signal source on X Series USB BNC devices, move the switch
under the BNC connector to the GS position.
With this type of connection, the NI-PGIA rejects both the common-mode noise in the signal and
the ground potential difference between the signal source and the device ground, shown as Vcm
in the figure.
AI+ and AI- must both remain within ±11 V of AI GND.
MIO X Series Device Configured in Differential Mode
PGIA
–
+
–
+
–
+
–
+
V
cm
V
s
Ground-
Referenced
Signal
Source
Common-
Mode
Noise and
Ground
Potential
AI GND
AI SENSE
Input Multiplexers
V
m
Measured
Voltage
Instrumentation
Amplifier
AI+
AI–
I/O Connector

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments 636 Series and is the answer not in the manual?

National Instruments 636 Series Specifications

General IconGeneral
BrandNational Instruments
Model636 Series
CategoryData Loggers
LanguageEnglish

Related product manuals