© National Instruments | 4-11
X Series User Manual
Table 4-3. MIO X Series Analog Input Configuration
AI Ground-Reference
Setting
*
Floating Signal Sources
(Not Connected to
Building Ground)
Ground-Referenced
Signal Sources
†
Examples:
• Ungrounded thermocouples
• Signal conditioning with
isolated outputs
• Battery devices
Example:
• Plug-in instruments with
non-isolated outputs
Differential
Non-Referenced
Single-Ended (NRSE)
Referenced Single-Ended
(RSE)
*
Refer to the Analog Input Ground-Reference Settings section for descriptions of the RSE, NRSE, and
DIFF modes and software considerations.
†
Refer to the
Connecting Ground-Referenced Signal Sources section for more information.
+
–
+
–
AI+
AI–
AI GND
Signal Source DAQ Device
+
–
+
–
AI+
AI–
AI GND
Signal Source DAQ Device
+
–
+
–
AI
AI
SENSE
AI GND
Signal Source DAQ Device
+
–
+
–
AI
AI GND
Signal Source DAQ Device
Ground-loop potential (V
A
– V
B
) are added
to measured signal.
NOT RECOMMENDED
+
–
+
–
AI
AI GND
V
B
V
A
Signal Source DAQ Device

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments 636 Series and is the answer not in the manual?

National Instruments 636 Series Specifications

General IconGeneral
BrandNational Instruments
Model636 Series
CategoryData Loggers
LanguageEnglish

Related product manuals