4-4 | ni.com
Chapter 4 Analog Input
Analog Input Ground-Reference Settings
MIO X Series devices support the following analog input ground-reference settings:
• Differential mode—In DIFF mode, the MIO X Series device measures the difference in
voltage between two AI signals.
• Referenced single-ended mode—In RSE mode, the MIO X Series device measures the
voltage of an AI signal relative to AI GND.
• Non-referenced single-ended mode—In NRSE mode, the MIO X Series device measures
the voltage of an AI signal relative to one of the AI SENSE inputs specific for that channel.
The AI ground-reference setting determines how you should connect your AI signals to the MIO
X Series device. Refer to the
Connecting Analog Input Signals section for more information.
Ground-reference settings are programmed on a per-channel basis. For example, you might
configure the device to scan 12 channels—four differentially-configured channels and eight
single-ended channels.
MIO X Series devices implement the different analog input ground-reference settings by routing
different signals to the NI-PGIA. The NI-PGIA is a differential amplifier. That is, the NI-PGIA
amplifies (or attenuates) the difference in voltage between its two inputs. The NI-PGIA drives
the ADC with this amplified voltage. The amount of amplification (the gain), is determined by
the analog input range, as shown in Figure 4-2.
Figure 4-2. MIO X Series Device NI-PGIA
V
in+
V
m
= [V
in+
– V
in–
] × Gain
V
m
V
in–
PGIA
+
–
Measured
Voltage

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments 636 Series and is the answer not in the manual?

National Instruments 636 Series Specifications

General IconGeneral
BrandNational Instruments
Model636 Series
CategoryData Loggers
LanguageEnglish

Related product manuals