48
Quantity of sample can be increased by adding on pan next details. Value of
pieces quantity and mass will be automatically increased and corrected by
balance software mass of single detail (SMP).
In balance software are four conditions of Automatic Accuracy Correction
function operating:
1. quantity of pieces must be increased
2. quantity of pieces after adding can not be bigger than double quantity
of previous pieces
3. current quantity of pieces must be in tolerance range ± 0,3 to total
value
4. result must be stable
If user finds quantity of pieces sufficient, it is necessary to save mass of single
detail to balance memory.
18.2.3. Wprowadzenie masy detalu z bazy danych wagi
After introducing assortment from assortment base sample mass SMP of
assortment is automatically introduced.
Press F8 key list of assortment will occur. Using directional keys (arrow up,
arrow down) choose required assortment. Press ENTER. Balance will return to
counting pieces and in left upper corner of display number of chosen
assortment and name of this assortment is displayed.
Fig. 33. Introducing mass of detail from assorment database
1- number of product from assortment database
2- Name of product
NOTICE:
For each assortment there must be introduced mass. This can be performed by
edition of assortment base.