Parameters:
<Position> <numeric value> (integer)
Range: 0 to 10
*RST: 10
Example: //Set the reference position to the 5th grid line
DISP:TRAC:Y:SPAC LOG
DISP:TRAC:Y:RPOS 5
Manual operation: See "Range / Ref Position" on page 370
See "Range / Ref Position" on page 380
INPut:ATTenuation
<Attenuation>
This command selects the RF attenuation.
Effects of this command
Turns off automatic selection of the attenuation and therefore the coupling to the
reference level (INPut:ATTenuation:AUTO).
Reference level will be adjusted if the currently defined reference level is not com-
patible to the selected attenuation.
Parameters:
<Attenuation> <numeric value>
Range: 0 to 40
Increment: 5
*RST: 0 (AUTO is ON)
Default unit: dB
Example: //Select attenuation
INP:ATT 30dB
Manual operation: See "Attenuator" on page 370
See "Attenuator" on page 381
See "Attenuator" on page 389
See "Attenuator" on page 393
INPut:ATTenuation:AUTO <State>
This command turns automatic selection of the attenuation level on and off.
Parameters:
<State> ON | OFF | 1 | 0
*RST: ON | 1
Example: //Turn on automatic attenuation selection
INP:ATT:AUTO ON
Manual operation: See "Attenuator" on page 370
See "Attenuator" on page 381
See "Attenuator" on page 389
See "Attenuator" on page 393
Remote Commands of the Spectrum Application

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FPC Series and is the answer not in the manual?

R&S FPC Series Specifications

General IconGeneral
BrandR&S
ModelFPC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals