RIGOL Chapter 2 Command System
2-138 MSO7000/DS7000 Programming Guide
[:SOURce[<n>]]:FREQuency[:FIXed]
[:SOURce[<n>]]:FREQuency[:FIXed] <frequency>
[:SOURce[<n>]]:FREQuency[:FIXed]?
If modulation is not enabled for the specified function/arbitrary waveform generator
channel, this command is used to set or query the output frequency of the specified
function/arbitrary waveform generator channel. If modulation is enabled for the
specified function/arbitrary waveform generator channel, this command is used to set
or query the carrier frequency of the specified function/arbitrary waveform generator
channel. By default, the unit is Hz.
When omitted, by default, the
operations are carried out on GI.
Sine: 100 mHz to
25 MHz
Square: 100 mHz
to 15 MHz
Pulse: 100 mHz to
1 MHz
Ramp: 100 mHz to
100 kHz
Arb: 100 mHz to
10 MHz
The query returns the frequency value in scientific notation. For example
2.0000000E+5.
:FREQuency 1000 /*Sets the output frequency of GI to 1 kHz.*/
:FREQuency? /*The query returns 1.000000E+3.*/
[:SOURce[<n>]]:PHASe[:ADJust]
[:SOURce[<n>]]:PHASe[:ADJust] <phase>
[:SOURce[<n>]]:PHASe[:ADJust]?
Sets or query the start phase of the signal of the specified function/arbitrary waveform
generator channel. By default, the unit is °.
When omitted, by default, the
operations are carried out on GI.
The query returns the start phase in scientific notation.
:PHASe 90 /*Sets the start phase of GI to 90°.*/
:PHASe? /*The query returns 9.000000E+1.*/