RIGOL Chapter 2 Command System
2-140 MSO7000/DS7000 Programming Guide
[:SOURce[<n>]]:FUNCtion:RAMP:SYMMetry
[:SOURce[<n>]]:FUNCtion:RAMP:SYMMetry <val>
[:SOURce[<n>]]:FUNCtion:RAMP:SYMMetry?
Sets or queries the symmetry of the Ramp waveforms output from the specified
function/arbitrary waveform generator channel.
When omitted, by default, the
operations are carried out on GI.
Symmetry is defined as the percentage that the rising period of the ramp takes up in
the whole period.
The query returns the current symmetry in scientific notation.
:FUNCtion:RAMP:SYMMetry 50 /*Sets the symmetry of Ramp waveform output
from GI to 50%.*/
:FUNCtion:RAMP:SYMMetry? /*The query returns 5.000000E+1.*/
[:SOURce[<n>]]:VOLTage[:LEVel][:IMMediate][:AMPLitude]
[:SOURce[<n>]]:VOLTage[:LEVel][:IMMediate][:AMPLitude] <amplitude>
[:SOURce[<n>]]:VOLTage[:LEVel][:IMMediate][:AMPLitude]?
Sets or query the amplitude of the signal output from the specified function/arbitrary
waveform generator channel. By default, the unit is Vpp.
When omitted, by default,
the operations are carried
out on GI.
Related to the current
output impedance
HighZ: 20 mVpp to 5
Vpp
50Ω: 10 mVpp to 2.5
Vpp
Send the [:SOURce[<n>]]:OUTPut[<n>]:IMPedance command to set the input
impedance.
The query returns the amplitude value in scientific notation.
:VOLTage 2 /*Sets the output amplitude of GI to 2 V.*/
:VOLTage? /*The query returns 2.0000000E0.*/