Safety instructions
CAUTION
Uncontrolled system movements and unintended radiation
exposure!
Injury to the patient and personnel, and radiation damage.
◆ Always verify that the patient is correctly positioned.
◆ Always observe the patient during system movements.
◆ Press a STOP key in any of the following situations:
• The patient is not positioned correctly during system
movements
• At any unintentional system movement (especially at
autorange)
• The patient table moves in a wrong direction
• The patient table does not stop as expected
• A key sticks or a movement does not stop immediately
when a key is released
• The Suspend key does not respond during a scan
◆ Press an EMERGENCY OFF key if the system does not respond
to the STOP keys in any hazardous situation.
◆ Shut down the system immediately if system malfunctions
are detected and notify the Customer Service at Siemens
Healthineers.