In case of emergency
In addition to the STOP keys at the CT unit, EMERGENCY OFF keys to
be provided by the customer have to be easily visible and within easy
reach. Country-specific regulations must be observed.
In extreme cases, you can press an EMERGENCY OFF key to shut
down the whole system. Radiation and system movements will be
stopped.
You must only press the EMERGENCY OFF key:
If the STOP key on the CT system does not respond.
If situations arise that could cause injury to the patient or user or
damage to the equipment (for example, liquids or particles getting
into the equipment, damage of cover parts, fire).
Only after all causes of danger have been found and remedied, the
unit may be switched on again by a controlled restart.
The EMERGENCY OFF key interrupts the power supply of the system.
Data can be lost.
Restarting after emergency stop
In case the system has been switched off by pressing the emergency
key in the examination room (for example, by accident), a controlled
restart has to be done.
All causes of danger have been found and remedied.
1 Wait for approximately 1 minute after power switch off.
2 Release the EMERGENCY OFF key provided by the customer.
3 Switch the system on again.
4 Perform the usual checkup procedure to calibrate the system.
2
2.1
In case of emergency 2
SOMATOM Scope |
syngo
CT VC50
Print No. HC-C2-015-G.626.08.01.02
19