Raptor
12
Danger When conducting resistance tests, be sure that the
circuit is earthed to some point. If it is a switch, one of the
sides must be connected to earth and the switch must be
closed.
Danger Before injecting current on the primary of a current
transformer, be sure that its secondary is closed. Otherwise,
high voltages may appear.
Danger Never operate with the system if you observe
severe damage to it or humidity on it.
Hazardous situations for the Raptor system
Caution In systems with Raptor-
SL units, when preparing to
conduct injections by pass-through turns, be sure that all the units
are powered.
This is necessary so that the thermal protection
systems work correctly.
Caution Do not inject using the pass-through turn simultaneously
with the auxiliary output.
Caution Do not try to lift the equipment using the fold-down
handle. Instead, use the top handle.