Substituting
the given values:
Risetime
= (7.50 - 2.50) X 1
µs
= 5 ms.
MIX {BASIC 466/464)
For
the
MIX
mode
of
operation,
the
same general
procedures
can be used. With
the
first
part
of
the
display
at
a sweep rate set
by
the
A
TIME/DIV
switch
and
the
second
part
of
the
display
at a sweep rate set
by
the B Tl
ME/DIV
switch, it is
not
necessary
to
switch
display
modes
to
ensure location
of
the
correct
pulse.
However, inaccuracies are
introduced
into
the
measurement
by
the transition from A
to
B sweeps. The B
DL
Y'D
mode
is the
most
accurate
and therefore
recommended mode
of
making
differential
time
measurements.
@
466/464 DM44 Operators