OPERATOR'S SPECIFICATIONS
Refer
to
the
service
manual
for
complete
specifications.
Specifications
given are
for
an
operating
range
of
C
to
+40°C
unless
otherwise
stated.
466/464 OSCILLOSCOPE
STORAGE
Stored Writing Speed
VAR PERS Full Scan Speed: 0.5
div/µs
over
the
center6
X 8
divisions.
It
is necessary
to
use
the
STORAGE
LEVEL
control
to
compromise
between
writing
speed and view
time.
466 and 464.
FAST
Full Scan Speed:
11
O
div/µs
over
the
center
6 X 8
divisions
(0.9
cm/div).
464
only.
FAST Full Scan Speed: 150
div/µs
over
the
center
6 X 8
divisions
(0.9
cm/div).
466
only.
FAST Reduced Scan Speed: 3,000
div/µs
over
the
center
8 X 10
divisions
(0.45
cm/div).
466
only.
VAR PERS Reduced Scan Speed: 3
div/µs
over the
center
8 X 10
divisions.
It is necessary
to
use the
STORAGE
LEVEL
control
to
compromise
between
writing
speed and view time. 466
only.
Storage View Time
FAST and VAR PERS View Times:
At
least 15 seconds.
These times are at full stored
display
intensity.
They
are
extended at least
25
times
by
using
reduced
intensity
in
the
SAVE Display mode.
VERTICAL
Deflection Factor Accuracy:
Within
3% in
the
calibrated
position.
Frequency Response: De
to
at least 100 MHz
for
CH 1
and
CH
2.
Ac-coupled,
low-frequency
response is 10 Hz
or
less. Use
of
a 10X
probe
extends
frequency
response
to
1 Hz.
@
466/464
DM44
Operators
65
..
J