TM 11-6625-2735-14-1
HORIZONTAL SYSTEM CHECK
Equipment Required
4. 500 MHz Filter
1. Time-Mark Generator
5. 42-inch 50-Ohm BNC Cable (2 required)
2. Sine-Wave Generator
6. 50-Ohm BNC Termination (2 required)
3. Amplitude Calibrator
7. 10X Probe (see Table 5-1, 10X Probe examples)
Control Settings
POWER
ON
Display
INTENSITY
Midrange
FOCUS
Midrange
SCALE ILLUM
Midrange
Vertical (CH 1 and CH 2)
POSITION
Midrange
VOLTS/DIV
0.5 V (new setting)
VAR VOLTS/DIV Calibrated detent
AC-GND-DC
DC
INVERT
Off (button out)
VERT MODE
CH 1
100 OR 20 MHz BW
Full bandwidth (push in,
then release)
Trigger (A and B)
COUPLING
AC
LEVEL
0
SLOPE
+
A TRIGGER SOURCE
NORM (new setting)
B TRIGGER SOURCE
STARTS AFTER DELAY
(new setting)
TRIG MODE
AUTO
A TRIGGER HOLDOFF
NORM
Sweep
A TIME/DIV
1 ms
B TIME/DIV
1 ms
VAR TIME/DIV Calibrated detent
DELAY TIME POSITION
Fully counterclockwise
HORIZ DISPLAY
A
X10 MAG
Off (button out)
POSITION (Horiz)
Midrange
FINE
Midrange
20. Check VAR TIME/DIV Range
a. Set the time-mark generator for 5 millisecond time
marks and connect through a 42-inch 50-ohm BNC cable,
to a 50-ohm BNC termination, to CH 1 input.
b. Set A TRIGGER LEVEL for a stable triggered
display, then set VAR TIME/DIV fully counterclockwise.
c. CHECK—That VAR TIME/DIV UNCAL light is il-
luminated.
d. CHECK—For at least 6 time marks between first and
last graticule lines.
e. Return VAR TIME/DIV control to calibrated detent.
5-15