TM 11-6625-2735-14-1
Characteristics
Temperature
Operating (AC)
Storage
Altitude
Operating
Storage
Humidity (Operating and Stor-
age)
Vibration (Operating)
Shock (Operating and Non-
operating)
Transportation
TABLE 1-2
ENVIRONMENTAL
Performance Requirements
–15°C to +55°C.
–55°C to +75°C.
To 15,000 feet. Maximum operating
temperature decreased 1°C/1,000 feet
above 5,000 feet.
To 50,000 feet.
5 cycles (120 hours) referenced to
MIL-E-16400F.
15 minutes along each of three major
axes at a total displacement of
0.025 inch P-P (4 g’s at 55 Hz) with
frequency varied from 10 Hz-to 55 Hz
to 10 Hz in one minute sweeps. After
sweep vibration in each axis, hold
frequency steady at each major res-
onance for 3 minutes, or if no such
resonances are found, hold at 55 Hz
for three minutes.
30 g’s, 1/2 sine, 11 ms duration, 2
shocks per axis each direction for
a total of 12 shocks.
Meets the limits of National Safe
Transit Committee test procedure 1A
with a 30-inch drop.
Supptemental Information
1-11