Troubleshooting
AFG3000 Series Arbitrary/Function Generators Service Manual 6-27
Figure 6-21: Generator troubleshooting procedure
Ye s
No
Execute the Utility >
Diagnostics.
AFG302xB?
Ye s
No
Ye s
No
Does the Trigger
Input/Output function?
Ye s
No
Do the
Ext Modulation Input,
Ext Ref Input, and
Ext Ref Output
function?
Ye s
No
Replace the CPU board.
1
Completed a Generator check
Generator troubleshoot start
Ye s
No
Diagnostics OK?
Is it an error code
related to CPU?
Replace the Generator board.
Execute the Utility >
Calibration.
Calibration OK?
Replace the Generator board.
Remove the Output Amplifier
board.
Check the signal of the
Pre-Amplifier.
Is the signal OK?
Replace the Generator board.
Replace the Amplifier
board.
Replace the Generator board.
Replace the Generator board.
Does
the Add Input function?
Replace the Amplifier board.
Ye s
No
No
Ye s