Performance Verification
4-26 AFG3000 Series Arbitrary/Function Generators Service Manual
AC Flatness Test
This test verifies the flatness of a sine wave to 100 kHz sine wave.
1. Connect the arbitrary/function generator to the power meter with a power head
as shown in Figure 4-6.
Figure 4-6: AC flatness tests
2. Set up the arbitrary/function generator using the following steps:
3. Set the frequency of the Power Meter to 100 kHz.
4. Write the Power Meter reading of the 100 kHz sine wave as a reference power
value.
5. Verify that the power measurement at each frequency is within the error
specified in the following tables.
Select menu Setting Operation
Function Sine Sine (front)
Frequency 100.0000 kHz Frequency/Period (front)
Amplitude Units dBm TopMenu > Amplitude/Level Menu > -more- >
Units > dBm
Amplitude + 4.0 dBm Amplitude/High (front)
Channel 1 Output On On (front)
AFG3011 Measurement Range
Function Amplitude Frequency (dB) (dB)
Sine + 4.0 dBm 100.00 kHz = Reference - - - - - -
Sine + 4.0 dBm 500.00 kHz Reference ± 0.15
Sine + 4.0 dBm 1.00 MHz Reference ± 0.15
Sine + 4.0 dBm 5.00 MHz Reference ± 0.30
Sine + 4.0 dBm 10.00 MHz Reference ± 0.30
Power meter
50 Ω BNC coaxial cable
CH1 Output
AFG3000 series
Power head

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix AFG3021 and is the answer not in the manual?

Tektronix AFG3021 Specifications

General IconGeneral
BrandTektronix
ModelAFG3021
CategoryPortable Generator
LanguageEnglish

Related product manuals