Performance Verification
AFG3000 Series Arbitrary/Function Generators Service Manual 4-27
6. (AFG3xx2 only) Repeat steps 3 through 5 for the channel 2 output.
AFG302xB Measurement Range
Function Amplitude Frequency (dB) (dB)
Sine + 4.0 dBm 100.00 kHz = Reference - - - - - -
Sine + 4.0 dBm 500.00 kHz Reference ± 0.15
Sine + 4.0 dBm 1.00 MHz Reference ± 0.15
Sine + 4.0 dBm 5.00 MHz Reference ± 0.30
Sine + 4.0 dBm 15.00 MHz Reference ± 0.30
Sine + 4.0 dBm 25.00 MHz Reference ± 0.50
AFG310x Measurement Range
Function Amplitude Frequency (dB) (dB)
Sine + 4.0 dBm 100.00 kHz = Reference - - - - - -
Sine + 4.0 dBm 1.00 MHz Reference ± 0.15
Sine + 4.0 dBm 5.00 MHz Reference ± 0.30
Sine + 4.0 dBm 25.00 MHz Reference ± 0.50
Sine + 4.0 dBm 50.00 MHz Reference ± 0.50
Sine + 4.0 dBm 100.00 MHz Reference ± 0.50
AFG325x Measurement Range
Function Amplitude Frequency (dB) (dB)
Sine + 4.0 dBm 100.00 kHz = Reference - - - - - -
Sine + 4.0 dBm 1.00 MHz Reference ± 0.15
Sine + 4.0 dBm 5.00 MHz Reference ± 0.30
Sine + 4.0 dBm 25.00 MHz Reference ± 0.50
Sine + 4.0 dBm 50.00 MHz Reference ± 0.50
Sine + 4.0 dBm 100.00 MHz Reference ± 1.00
Sine + 4.0 dBm 150.00 MHz Reference ± 1.00
Sine + 4.0 dBm 240.00 MHz Reference ± 2.00