Window trigger minimum timing
requirements
≥4 GHz models
Minimum
pulse width
Minimum
rearm time
Setup time Hold time
Window enters 150 ps 500 ps N.A. N.A.
Window exits 500 ps 150 ps N.A. N.A.
Logic qualified window enters 150 ps 500 ps N.A. N.A.
Logic qualified window exits 500 ps 150 ps N.A. N.A.
Window inside >t N.A. 500 ps N.A. N.A.
Window outside >t N.A. 150 ps N.A. N.A.
<4 GHz models
Minimum
pulse width
Minimum
rearm time
Setup time Hold time
Window 225 ps 300 ps N.A. N.A.
Logic qualified window 190 ps 300 ps -20 ps 0 ps
Transition trigger, minimum timing
requirements
The transition trigger event occurs at the end of the transition. The logic condition must be valid at
the moment the transition event occurs minus the setup time plus the hold time. Rearm time is the
time below the lower threshold and the time above the upper threshold. Neither time can be
violated.
≥4 GHz models
Minimum
pulse width
Minimum
rearm time
Setup time Hold time
Transition type 0 ps 500 ps N.A. N.A.
Logic qualified transition type 0 ps 500 ps 50 ps 50 ps
<4 GHz models
Minimum
pulse width
Minimum
rearm time
Setup time Hold time
Transition type 0 ps 225 ps N.A. N.A.
Logic qualified transition type 330 ps 225 ps 225 ps 175 ps
Transition trigger delta time range 1 ns to 1 s
Setup-Hold violation trigger, setup
and hold time ranges
Minimum Maximum
Setup time -100 ns 100 ns
Hold time -1 ns 102 ns
Setup + hold time 500 ps
Trigger specifications (cont.) Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series 63