Model 2601B-PULSE System SourceMeter® Instrument User's Manual Section 1: Introduction
2601B-PULSE-900-01 Rev. A August 2021 1-3
Power and environmental ratings
The 2601B-PULSE power and environmental ratings and connections are listed in the
following tables.
Category
Specification
Power supply
100 V ac to 240 V ac, 50 Hz or 60 Hz (autosensing). 240 VA maximum
Input and output connections
See Front panel (on page 3-1) and Rear panel (on page 3-5)
Environmental conditions
For indoor use only.
Operating altitude: Maximum 2000 meters (6562 feet) above sea level
Operating temperature: 0 °C to 35 °C at up to 70% relative humidity; at
35 °C to 50 °C, derate 3% relative humidity per °C
Storage: 25 °C to 65 °C
Pollution degree: 1 or 2
Source output electrical ratings
Voltage
40 V dc maximum
Current
3 A maximum at 6 V dc, 1 A maximum at 40 V dc
SMU pulse output, region 4
Region maximums
10 A at 20 V
Maximum pulse width
1.8 ms
Maximum duty cycle
1%
Measure input electrical ratings
Measurement category
O
Voltage
40 V dc maximum HI to LO
Current
3 A maximum at 6 V dc, 1 A maximum at 40 V dc
Impedance
Variable

Table of Contents

Other manuals for Tektronix Keithley 2601B-PULSE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 2601B-PULSE and is the answer not in the manual?

Tektronix Keithley 2601B-PULSE Specifications

General IconGeneral
BrandTektronix
ModelKeithley 2601B-PULSE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals