Settings ignored when the pulser is disabled
The following pulser settings do not affect source-measure unit (SMU) operation when the pulser
is disabled:
• smua.pulser.measure.aperture
• smua.pulser.measure.delay
• smua.pulser.protect.sensev
• smua.pulser.protect.sourcev
• smua.pulser.rangeY
Examples
The following topics contain TSP programming examples for the 2601B-PULSE pulser. You can use
these examples as a starting point for your own applications, editing them as needed.
Example 1: Ten-point pulse sweep
The code in this example uses the trigger model to program a 10-point pulse sweep of 10 µs pulses
from 1 A to 10 A with a 1 ms pulse period. A 1 Ω resistor is used as the device under test (DUT). The
following figure illustrates the pulse action in the example.
Figure 53: 10-point pulse sweep