Time Measurements panel
Use the T
ime Measurements panel to add timing-related measurements to the Results bar. Timing measurements can be taken on
time-domain analog, math, and reference waveforms. Timing measurements can also be taken on some digital channel signals.
To open the Time Measurements panel:
1. Tap the Measure button.
2. Tap the Time Measurements panel.
To add a measurement to the Results bar:
1. Select the signal source.
2. Select a measurement. If a measurement requires two signal sources, select both sources.
3. Tap Add.
Time Measurements panel
Measurement Description
Period
4
The time between two adjacent crossings of the Mid reference level (one cycle) of the waveform.
The measurement is taken on each cycle of the waveform record or measurement region.
Frequency
4
The frequency of the waveform. Frequency is the reciprocal of Period (Frequency = 1/Period).
Delay
The time difference between the mid reference (default 50%) amplitude point of two different
waveforms. Y
ou can specify the signal edges to measure in the measurement’s Configuration menu.
The measurement requires two sources.
Phase
4
The time difference (phase shift) between the specified signal edges of waveform source 1 and
waveform source 2. The measurement is expressed in degrees, where 360° comprise one waveform
cycle. This measurement requires two sources.
The measurement is taken on each cycle of the waveform record.
Positive Pulse Width
4
The distance (time) between the mid reference (default 50%) amplitude points of a positive pulse.
The measurement is made on the first pulse in the measurement region.
Negative Pulse Width
4
The distance (time) between the mid reference (default 50%) amplitude points of a negative pulse.
The measurement is taken on each cycle in the waveform record or measurement region.
Positive Duty Cycle
4
The ratio of the positive pulse width to the signal period, expressed as a percentage.
The duty cycle is measured on the first cycle in the measurement region.
Negative Duty Cycle
4
The ratio of the negative pulse width to the signal period, expressed as a percentage.
The duty cycle is measured on the first cycle in the measurement region.
Table continued…
4
This measurement can also be taken on digital signals.
Menus and dialog boxes
3 Series Mixed Domain Oscilloscope Printable Help 104

Table of Contents

Other manuals for Tektronix MDO32

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO32 and is the answer not in the manual?

Tektronix MDO32 Specifications

General IconGeneral
BrandTektronix
ModelMDO32
CategoryTest Equipment
LanguageEnglish

Related product manuals