Fields or controls Description
Spectrogram Toggles the spectrogram traces on or off. The spectrogram display is useful for monitoring slowly-
changing RF phenomena. The x-axis represents frequency
, just like the typical spectrum display. The
y-axis represents time. Color indicates amplitude.
Spectrogram slices are generated by taking each spectrum and flipping it on its edge so that it is one
pixel row tall, and then assigning colors to each pixel based on the amplitude at that frequency with
the cold colors of blue and green representing low amplitude, and the hotter colors of yellow and red
indicating higher amplitude. Each new acquisition adds another slice at the bottom of the spectrogram,
and the history moves up one row. See Spectrogram display on page 244 for more information.
Graticule Style Sets the type of graticule to display.
Full displays a frame, cross hairs, and a grid on the display. This style is useful for making quick,
full-screen measurements with cursors and automatic readouts when cross hairs are not needed.
Grid, Solid, and Cross Hair provide compromises between Frame and Full.
Frame provides a clean screen on which you can most easily read automatic measurement results
and other screen text.
Graticule Intensity Sets the brightness of the graticule. Tap the field and use the A knob to set the graticule intensity.
Screen Annotation Double-tap the field and use the keyboard to add text to the display.
Selected Trace Makes the Normal, Average, Max Hold, or Min Hold trace the selected trace.
Cursors menu (RF view)
Use the RF view cursors to take manual measurements on spectrum signals.
T
o open the RF view cursors menu, double-tap on a cursor or its readout in the Waveform View window.
Cursors menu (RF view), fields and controls
Field or control Description
Display Toggles the spectrum cursor display On or Off.
Reference To Center Moves the reference cursor to the center of the display.
Cursor A X-Position Sets a specific x-axis position for Cursor A using the multipurpose knob or the virtual keypad.
Cursor B X-Position Sets a specific x-axis position for Cursor B using the multipurpose knob or the virtual keypad.
Readout Sets the cursor readouts to show Absolute or Delta values.
Delta readouts are relative to the Reference Marker
.
Menus and dialog boxes
3 Series Mixed Domain Oscilloscope Printable Help 232

Table of Contents

Other manuals for Tektronix MDO32

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO32 and is the answer not in the manual?

Tektronix MDO32 Specifications

General IconGeneral
BrandTektronix
ModelMDO32
CategoryTest Equipment
LanguageEnglish

Related product manuals