Acquiring Waveforms
3-52
CSA7000 Series, TDS7000 Series, & TDS6000 S eries Instruments User Manual
Overview Control elements and resourcesTo turn Fast Acquisitions on and off (Cont.)
To set the
style
6. Touch the DISP button and select the Appearance tab.
7. Select between Vectors, Dots, or Inten Samp display
styles. (Dots is the default setti ng when in Equivalent
Time (ET) acquisition mode. Vectors is the default when
not in ET acquisition mode.)
8. Select between Off, Variable, and Infinite Display
Persistence. (Off is the factory default setting.)
Variable persistence is explained on page 3--140. No
Persistence erases the display each time a new
waveform array is displayed.
If your instrument supports mask testing, Mask Autofit
can affect persistence. When enabled Autofit moves the
waveform vertically and horizontally to reduce the
number of hits within a given mask. If infinite or variable
persistence is enabled, the movements will clear any
persisted data. If Autofit makes frequent adjustments,
there can be very little or even no persisted data
displayed.
9. If you select Variable, touch Persist Time and adjust the
persistence time (decay rate).

Table of Contents

Other manuals for Tektronix TDS6000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS6000 Series and is the answer not in the manual?

Tektronix TDS6000 Series Specifications

General IconGeneral
BrandTektronix
ModelTDS6000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals