Displaying Waveforms
CSA7000 Series, TDS7000 Series, & TDS6000 S eries Instruments User Manual
3- 127
Table 3- 7: Operations performed based on the waveform type (Cont.)
Control function Operating notesWaveform supports
Ch Ref
1
Math
Horizontal Scale Yes Yes Yes Wavef orms are adjusted according to the Zoom Lock setting.
Horizontal Position Yes Yes Yes
Horizontal Record Length Yes No No
Quick Horizontal and Vertical
Scale Adjust (Zoom)
Yes Yes Yes Dragging a box around a portion of the selected waveform adjusts
horizontal scale to fill the zoom graticule with the boxed portion (see
Setting MultiView Zoom Controls on page 3--131).
1
Pixel-map reference waveforms, those saved while in fast acquisitions mode or waveform database mode, cannot be
repositioned or rescaled.
Graticules. Select a graticule size from the Graticule Size drop down list in the
Zoom Setup window to change the size of the acquisition waveform and zoomed
waveform windows. The 50-50 button allocates half of the available display for
the zoomed graticule and half of the available display for the acquisition
window. The 80-20 button allocates 80% of the available display for the zoomed
graticule and 20% for the acquisition window. Touch 100 to use the entire
display for the zoomed graticule.
Figure 3--31 on page 3--124 shows the elements of the graticules; the elements
are the same for each graticule displayed.
Operations on the Timebase. In general, the method of adjusting (horizontally
scaling, setting resolution/record length, positioning, and so on) is from the front
panel: adjust the timebase using the Horizontal Scale, Resolution, and Position
knobs. Only channel waveforms can be set directly.
Table 3--7 shows how horizontal operations relate to the waveform types; the key
points follow:
H The instrument displays a reference waveform with horizontal settings in
effect at the time it was saved. You cannot adjust these settings. See Saving
and Recalling Waveforms on page 3--253 for more information on reference
waveforms.

Table of Contents

Other manuals for Tektronix TDS6000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS6000 Series and is the answer not in the manual?

Tektronix TDS6000 Series Specifications

General IconGeneral
BrandTektronix
ModelTDS6000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals