Adjustment Procedures
5-10
TDS7104 & TDS7054 Service Manual
4. Locate the test program file: After the installation is complete, open the
directory C:\tekcats\ and then locate the file tds7000_nt.exe.Thisisthefile
you will double click to execute either the performance verification
procedure or the adjustment procedure. The install procedure also created a
desktop shortcut (link) to this file for your convenience.
Adjustment Program Execution
Once the instrumentation is connected, configured, and has been allowed to
warm up for at least 20 minutes, you can start the adjustment program. To start
the adjustment program, select the file C:\tekcats\tds7000_nt.exe (or the
desktop shortcut).
Program execution is a matter of following the on-screen instructions. The
program contains help screens to provide additional information you may need to
know to execute the steps. When the program asks whether you want to execute
the performance verification or the adjustment procedure, select the adjustment
procedure.

Table of Contents

Other manuals for Tektronix TDS7054

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS7054 and is the answer not in the manual?

Tektronix TDS7054 Specifications

General IconGeneral
BrandTektronix
ModelTDS7054
CategoryTest Equipment
LanguageEnglish

Related product manuals