Video and General Performance Verication
Equalized Eye Gain
(Options EYE or PHY)
This test check
s for the presence of an internally generated eye signal.
1. Press FACTORY.
2. HD only: Connect an HDVG1 output to the SDI A input and terminate the
loopthrough connectors. Set the HDVG1 to generate a 1080 59.94i 100%
Color Bars s
ignal.
3. SD only: Connect a DVG1 output to the SDI A input and terminate the
loopthroug
h connectors. Set the DVG1 to generate a 525-270 75% Color
Bars signal.
4. Press the E
YE button (or MEAS if no EYE button) to get the Eye display.
5. Press CONFIG, and then select Physical Layer Settings > Eye Display Mode
> Equaliz
ed.
6. Using the graticule, check that the waveform amplitude is between 400 mV
and 650 m
V.
7. Record Pass or fail in the test record.
92 Waveform Rasterizers Specications and Performance Verication