Table 33: Jitter Display characteristics (Option PHY) (cont.)
Characteristic Performance requirement Reference information
Resolution 0.01 UI
Jitter Waveform
Amplitude Error, typical
Error for jitter frequencies between HPF
setting and 5x HPF setting due to loop
peaking, maximum of 1.5 dB.
HD at 10 Hz HPF
Within 0.4 UI + 10% of reading for jitter
frequencies from 5x bandwidth selection to
4MHz.
HD at 100 Hz to 100 kHz
HPF
SD at All HPF Settings
Within 0.1 UI + 10% of reading for jitter
frequencies from 5x bandwidth selection to
4MHz.
Scale
Horizontal 1 Line, 2 Line, 1 Field, and 2 Field modes.
Vertical
1 UI/div, 0.2 UI/div, and v ariable 0.1 UI/div
to 2.5 UI/div.
Jitter Measurement Time 0.5 s
Approximately 10,000,000 independent
samples.
Tab
le 34: Signal level and cable length detector characteristics
Characteristic Performance requirement Reference information
Source Level Detector
Me
asurement Accuracy,
typical
40
mV at 800 mV level, 0 to 100 m Belden
8281 cable.
Me
asurement Resolution,
typical
10
mV
M
easurement Dynamic
Range, typical
± 50% from 800 mV.
Cable Length Detector
Measurement Error
15 m 10% of the reading for the cable type. Requires clean SD signal with rise time less
than 1 ns for full accuracy. Excessive ringing
or slower rise time may appear as increased
cable length.
Measurement Resolution 2 m
Measurement Range, typical
Cable loss range of 0 to 40 dB at ½ serial
bit rate. Equivalent to:
HD: 0 to 120 m Belden 8281
SD: 0 to 400 m Belden 8281
Waveform Rasterizers Specications and Performance Verication 2 1

Table of Contents

Other manuals for Tektronix WVR6020

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix WVR6020 and is the answer not in the manual?

Tektronix WVR6020 Specifications

General IconGeneral
BrandTektronix
ModelWVR6020
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals