Test Records
Table 58: SD Eye Performance Test Record (WVR7120 with Options EYE or PHY) (cont.)
Performance Test Minimum Incoming Outgoing Maximum
SD Cable Meter
(100 m), SDIA
76 m 127 m
SD Cable M eter (0 m),
SDIB
0m 17m
SD Cable Meter
(100 m), SDIB
76 m 127 m
SD Jitter Readout
Noise
Pass
SD Jitter Waveform
Noise(optionPHY)
Pass
SD 10 second Jitter
Bounce
Pass
SD Jitter Readout
Accuracy
Pass
SD Jitter Waveform
Accuracy (option PHY)
Pass
SD Recovered Sine
Wave Fidelity
Pass
Video Performance Test Record (Option CPS)
Use the following table to record the results from the video performance test for
instruments with Option CPS installed.
Table 59: WVR6020 / WVR7020 / WVR7120 Waveform Rasterizer Video Performance Test Record (Option CPS)
Instrument Serial Number: Certificate Number:
Temperature:
RH %:
Date of Calibration:
Technician:
Performance Test Minimum Incoming Outgoing Maximum
Composite Analog
Vertical Measurement
Accuracy
99%
(707.71mv)
101%
(722mv)
Composite Analog
Frequency Response
99%
(–7mv)
101%
(7mv)
Waveform Rasterizers Specifications and Performance Verification 4 3