Test R ecords
Table 59: WVR6020 / WVR7020 / WVR7120 Waveform Rasterizer Video Performance Test Record (Option CPS) (cont.)
Performance Test Minimum Incoming Outgoing Maximum
Composite Analog
Delay Variation Over
Frequency (Group
Delay)
-10ns
(–7.1mv)
+10ns
(7.1mv)
Composite Analog
PulsetoBarRatio
99%
(–7mv)
101%
(7mv)
Composite Analog
Field Rate Tilt
–0.5%
(–3.5mv)
+0.5%
(+3.5mv)
Composite Analog
Line Rate Tilt
–0.5%
(–3.5mv)
+0.5%
(+3.5mv)
Composite Analog
Input DC Offset with
Restore Off
–10mv 10mv
Composite Analog
Input Clamp Off Check
Pass/Fail
Composite Analog Input DC Restore Hum Attenuation
Fast Mode
(0mv) 5%
(35mv)
Slow Mode 90%
(643mv)
110%
(785mv)
External Reference
Lock R ange
-50ppm 50ppm
VITC Decoding
Functionality
Pass
Audio Performance Test Record (Options AD and DDE)
Use the following table to record the results from the audio performance test for
instruments with Options AD or DDE installed.
Table 60: WVR6020 / WVR7020 / WVR7120 Waveform Rasterizer Audio Peformance Test Record (Options AD and DDE)
Instrument Serial Number: Certicate Number:
Temperature:
RH %:
Date of Calibration:
Technician:
Performance Test Minimum Incoming Outgoing M aximum
AES Audio Input Sample Rate
AESA1-2In
(96 kHz)
Pass
44 Waveform Rasterizers Specications and Performance Verication

Table of Contents

Other manuals for Tektronix WVR6020

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix WVR6020 and is the answer not in the manual?

Tektronix WVR6020 Specifications

General IconGeneral
BrandTektronix
ModelWVR6020
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals