Specications
This chapter contains specications for the Te ktronix WVR6020, WVR7020, and
WVR7120 waveform rasterizers. Items listed in the Performance Requirement
column are ge
nerally quantitative, and are either tested by the Performance
Verication procedure or are guaranteed by design.
Items listed in the Reference Information column are useful operating parameters
that have typical values; information in this column is not guaranteed.
The specications listed in the Electrical Specications portion of these tables
apply over an ambient temperature range of +0 °C to +50 °C. The rated accuracies
are valid when the instrument is calibrated in an ambient temperature range of
+20 °C to +30 °C and has been operating continuously for at least 20 minutes
within this range.
The characteristics are divided into the following sections:
Common Characteristics
High Denition (HD) Characteristics
Standard Denition (SD) Characteristics
Options Characteristics
Common Characteristics
The characteristics in this section are common to the WVR6020, WVR7020,
and WVR7120 Waveform Rasterizers.
Table 1: SDI input waveform vertical characteristics
Characteristic Performance requirement Reference information
YPbPr signal from a digital signal generator
as measured in YPbPr mode. Limited by
display resolution, measured by Graticules
or Cursors. Applies to all three channels.
At X1
±0.5%
Displayed
Vertical Gain
Accuracy
At X5
±0.2%
Variable Gain Range
0.25X to 10X
Line Rate Tilt
< 0.1%, guaranteed by digital design.
Off Screen Recovery < 0.1% variation in baseline of a 5 M Hz
modulated pulse when positioned anywhere
on screen.
X1, X5, or with any variable gain setting.
RGB Transcoder Accuracy ± 0.1%
Bit Integrity Accurately shows both 8 and 10 bit signals.
Waveform Rasterizers Specications and Performance Verication 1

Table of Contents

Other manuals for Tektronix WVR6020

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix WVR6020 and is the answer not in the manual?

Tektronix WVR6020 Specifications

General IconGeneral
BrandTektronix
ModelWVR6020
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals