Chapter 6: Replacing Assemblies
To remove and replace the power supply
6–33
To remove and replace the power supply
When necessary, refer to other removal procedures.
WARNING SHOCK HAZARD!
If the power supply is defective it could have a dangerous charge on some capacitors. This charge
could remain for many days after removing power from the supply.
1
Disconnect the power cable and remove the top cover.
2 Remove the floppy disk drive.
3 Remove the motherboard subassembly.
4 Remove the power supply spring cover.
5 Disconnect the cables from the front of the supply.
Figure 5-32
Disconnecting Power Supply Cables
Red wire
Green wire
Yellow wire
Brown wire
Power supply
spring cover
Jumper cable