EasyManuals Logo
Home>Agilent Technologies>Test Equipment>8156A

Agilent Technologies 8156A User Manual

Agilent Technologies 8156A
291 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #107 background imageLoading...
Page #107 background image
106
Remote Commands
INPut Commands
8.5 INPut Commands
:INPut:ATTenuation
Syntax :INPut:ATTenuation <wsp>
<value>[DB]|MIN|DEF|MAX
Description This command sets the attenuation factor for
the instrument. The attenuation factor is used,
with the calibration factor (see ) to set the filter
attenuation.
Attenuation
filter
(dB) = Att(dB) - Cal(dB)
You set the attenuation factor by sending a
value (default units are dB), or by sending
MIN, DEF or MAX, which specify the
minimum, default and maximum values for the
attenuation factor.
The minimum value and the default value are
those values for which Attenuation
filter
= 0dB.
The maximum value is that value for which
Attenuation
filter
is at its greatest.
:INPut:ATTenuation?
Syntax :INPut:ATTenuation? [<wsp>
MIN|DEF|MAX]
Description The query returns the current attenuation
factor, in dB.
Attenuation
filter
(dB) = Att(dB) - Cal(dB)
By sending MIN, DEF, or MAX with the query
the minimum, default or maximum value
possible for the attenuation factor is returned.
Example OUTPUT 728;":INP:ATT 32.15"
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8156A and is the answer not in the manual?

Agilent Technologies 8156A Specifications

General IconGeneral
BrandAgilent Technologies
Model8156A
CategoryTest Equipment
LanguageEnglish

Related product manuals