EasyManuals Logo
Home>Agilent Technologies>Test Equipment>8156A

Agilent Technologies 8156A User Manual

Agilent Technologies 8156A
291 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #114 background imageLoading...
Page #114 background image
113
Remote Commands
OUTPut Commands
:OUTPut:POWer?
Syntax :OUTPut:POWer? [<wsp> MIN|DEF|MAX]
Description The query returns the current through-power,
in dBm.
ThroughPower(dBm) = ThroughPower
Base
(dBm) + Att
filter@Base
(dB)-Att
filter
(dB)
By sending MIN, DEF, or MAX with the query
the minimum, default or maximum value
possible for the through-power is returned.
Example OUTPUT 728;":OUTP:POW 32.15"
OUTPUT 728;":OUTP:POW?"
ENTER 728;A$
:OUTPut:[:STATe]
Syntax :OUTPut[:STATe] <wsp> OFF|ON|0|1
Description This command sets the state of the output
shutter, that is, whether it is open or closed.
OFF or 0 closes the shutter, and no power gets
through. ON or 1 opens the shutter, and power
gets through.
:OUTPut[:STATe]?
Syntax :OUTPut[:STATe]?
Description The query returns whether the output shutter is
open or closed.
0 indicates the shutter is closed (no power is
getting through). 1 indicates that the shutter is
open (power is getting through).
Example OUTPUT 728;":OUTP ON"
OUTPUT 728;":OUTP?"
ENTER 728;A$
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8156A and is the answer not in the manual?

Agilent Technologies 8156A Specifications

General IconGeneral
BrandAgilent Technologies
Model8156A
CategoryTest Equipment
LanguageEnglish

Related product manuals