ABORt
ABORt
Halts a sequence, list, sweep, or burst, even an infinite burst. Also causes trigger subsystem to return to idle state. If
INITiate:CONTinuous is ON, instrument immediately proceeds to wait-for-trigger state.
Parameter Typical Return
(none) (none)
Halt the items listed above:
ABOR
l Halts any triggered action (triggered list, triggered sweep, triggered burst, triggered arbitrary waveform playback).
l ABORt has no effect when instrument is in normal or modulated modes, except for sequenced arbitrary wave-
forms, lists, bursts, and sweeps. If instrument is running a sequence, list, burst, or sweep, ABORt restarts the
stopped item with the current INIT and trigger conditions.
l When ABORt occurs in list mode, the frequency goes back to the "normal" mode frequency until the first trigger
occurs. After the first trigger, the first frequency in the list will be used.
l If ABORt executed during sweep, sweep returns to starting sweep frequency.
l ABORt always applies to both channels in a two-channel instrument.
156 Agilent 33500 Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 33500 Series and is the answer not in the manual?

Agilent Technologies Agilent 33500 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 33500 Series
CategoryInverter
LanguageEnglish

Related product manuals