DC Offset Voltage Verification
DC Offset Voltage Verification
Checks the DC Offset Voltage on two attenuator ranges:
1. Set the DMM to measure DCV. Connect the DMM to the channel output as shown below.
2. Set the instrument to each output in the table below and measure the output voltage with the DMM.
Use Waveform Generator Measurement
Output Setup Function Voltage Nominal Error*
Q High Z DC 0.0 V 0.0 VDC ±0.002 VDC
Q High Z DC 500 mV 0.500 VDC ±0.007 VDC
Q High Z DC 10.0 V 10.0 VDC ±0.102 VDC
* Based upon 1% of setting ±2 mVDC for High-Z.
3. Compare the measured value to the test limits shown in the table.
4. Two-channel instruments only: connect DMM to channel 2 output and repeat steps 2 and 3.
380 Agilent 33500 Series Operating and Service Guide