Configure a Ramp Wave
Configure a Ramp Wave
Description
A ramp wave has amplitude, offset, and phase relative to sync pulse. It also has symmetry for creating triangular and
other similar waveforms. Its amplitude and offset can also be set using high and low voltage values.
Example
The following waveform can be set up with the series of SCPI commands, where high and low can be used in place of
SOUR:VOLT and SOUR:VOLT:OFFS.
The following commands produce the ramp wave shown above.
FUNCtion RAMP
FUNCtion:RAMP:SYMMetry +50.0
FREQ +1.0E+03
VOLTage +2.0
VOLTage:HIGH +2.0
VOLTage:LOW +0.0
VOLTage:OFFSet +1.0
OUTPut1 1
Remarks
l Ramp frequency is limited to 200 kHz.
l Although period can be adjusted from the instrument's front panel, there is no SOUR:FUNC:RAMP:PER or
SOUR:PER command that can be used in addition to SOUR:FREQ.
330 Agilent 33500 Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 33500 Series and is the answer not in the manual?

Agilent Technologies Agilent 33500 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 33500 Series
CategoryInverter
LanguageEnglish

Related product manuals