Self Calibration Adjustment (Channel 2)
Self Calibration Adjustment (Channel 2)
1. Enter and begin the following setup.
Setup
54* Self-calibration. The output is disabled.
* Constants are stored after completing this setup.
2. After performing setup 54:
a. If your calibration procedures require you to verify the adjustment just made, exit the calibration menu and
perform DC Offset Voltage Verification. Be sure to do this for channel 2.
b. If you are making all of the adjustments and then verifying the instrument’s performance, continue with the
next procedure in this section.
This setup requires approximately 15 seconds to complete.
Agilent 33500 Series Operating and Service Guide 399