Self-Test
Self-Test
Run self-test to ensure that the instrument is in working order before beginning any additional adjustments.
Be sure to unlock the instrument and follow the requirements listed in Test Considerations before beginning any adjust-
ments.
1. Press System, then Instr Setup, then Calibrate. Enter setup number 1 and select BEGIN.
Setup
1 Performs the self-test. The Main Output is disabled during test.
2. If the instrument fails any self-test, you must repair the instrument before continuing the adjustment pro-
cedures.
A complete self-test (*TST?) takes approximately 15 seconds.
388 Agilent 33500 Series Operating and Service Guide