MARKer Subsystem
l When sweep is enabled, marker frequency must be between start frequency and stop frequency. Attempting to
set the marker cycle outside this range will set marker frequency to start frequency or frequency (whichever is
closer). From the remote interface, a "Settings conflict" error will also be generated.
[SOURce[1|2]:]MARKer:POINt {<sample_number>|MINimum|MAXimum}
[SOURce[1|2]:]MARKer:POINt? [{MINimum|MAXimum}]
Sets the sample number at which the front-panel Sync signal goes low within the active arbitrary waveform. OUT-
Put:SYNC:POLarity may reverse this..
Parameter Typical Return
Whole number from 4 to number of samples in waveform, minus 3; default is mid-
point of arbitrary waveform
+1.000000000000000E+01
Set marker point to 10th sample in waveform:
MARK:POIN 10
l Command only sets marker point in currently active arbitrary waveform (FUNCtion:ARBitrary), not in a sequence.
l Command is valid only under these conditions:
l OUTPut:SYNC:MODE set to MARK, FUNC set to ARB, FREQuency:MODE set to CW
l OUTPut:SYNC:MODE set to CARR, FUNC set to ARB, BURSt ON.
l OUTPut:SYNC:MODE set to CARR, FUNC set to ARB, FREQuency:MODE set to SWEEP
l OUTPut:SYNC:MODE set to MARK, internal modulation active, and either FUNCtion set to ARB or a modulating
waveform's internal function is set to ARB
l OUTPut:SYNC:MODE set to MARK, external modulation active, and FUNCtion set to ARB
Agilent 33500 Series Operating and Service Guide 245