VOLTage Subsystem Introduction
Parameter Typical Return
{OFF|0|ON|1|ONCE}, default ON 0ξ˜ƒ(OFF)ξ˜ƒorξ˜ƒ1ξ˜ƒ(ON)
Turn voltage autoranging OFF:
VOLT:RANG:AUTO 0
l In the default mode, autoranging is enabled and the instrument automatically selects the optimal settings for the
output waveform generator and attenuator.
l With autoranging disabled (OFF), the instrument uses the instrument's current gain and attenuator settings.
l The APPLy command overrides the voltage autorange setting and automatically enables autoranging (ON).
l Disabling autoranging eliminates momentary disruptions caused by attenuator switching while changing ampli-
tude. However, the amplitude and offset accuracy and resolution (and waveform fidelity) may be adversely
affected when reducing the amplitude below the expected range change.
l If a VOLTage:COUPle[:STATe] is ON, changing this setting on either channel changes it on both.
[SOURce[1|2]:]VOLTage:UNIT {VPP|VRMS|DBM}
[SOURce[1|2]:]VOLTage:UNIT?
Selects the units for output amplitude.
Parameter Typical Return
{VPP|VRMS|DBM}, default VPP VPP, VRMS, or DBM
Set output amplitude units to Vrms:
VOLT:UNIT VRMS
l Does not affect offset voltage (VOLTage:OFFSet), high level (VOLTage:HIGH) or low level (VOLTage:LOW). They all
use units of volts.
l The instrument uses the current units selection for both front panel and remote interface operations. For exam-
ple, if you select "VRMS" from the remote interface (VOLTage:UNIT VRMS), the units are displayed as "VRMS" on
the front panel.
l Command applies to VOLTage? query results.
l Output units for amplitude cannot be set to dBm if the output termination is set to "high impedance." The units
are automatically converted to Vpp.
l Arbitrary waveform sequences do not accept units of Vrms or dBm.
l Unless you specify the units as part of either the VOLTage command or one of the APPLy commands, the VOLT-
age:UNIT command takes precedence. For example, if you select VOLTage:UNIT VRMS and do not include units
with an APPLy command, the <amplitude> in the APPLy command will be in "Vrms".
326 Agilent 33500 Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 33500 Series and is the answer not in the manual?

Agilent Technologies Agilent 33500 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 33500 Series
CategoryInverter
LanguageEnglish

Related product manuals