Basic Digital Operation
Using the Dual ARB Waveform Player
2. Create the first waveform segment:
a. Press
Mode > Dual ARB > Waveform Segments > Load Store to Store.
b. Highlight the default segment AUTOGEN_WAVEFORM.
c. Press
Rename Segment > Editing Keys > Clear Text.
d. Enter a file name (for example, TTONE), and press Enter > Store Segment To NVWFM Memory.
This renames the waveform segment, and stores a copy in non volatile memory.
3. Generate the second waveform:
a. Press
Mode > Multitone > Initialize Table > Number Of Tones > 9 > Enter > Done.
b. Press Multitone Off On to On, then Off.
Remember that a waveform cannot be renamed as a segment while it is in use.
This generates a multitone waveform with nine tones. During waveform generation, the M–TONE
and I/Q annunciators activate. The waveform is stored in volatile memory with the default file
name AUTOGEN_WAVEFORM.
4. Create the second waveform segment:
Repeat Step 2, giving this segment a descriptive name (for example, MTONE).
Building and Storing a Waveform Sequence
This example shows how to build and edit a waveform sequence using the two waveform segments
created on page 92. To use a segment in a sequence, the segment must reside in volatile memory; for
information on loading waveform segments from non volatile to volatile memory, see page 95.
1. Select the waveform segments:
Define a sequence as one repetition of the two–tone waveform segment followed by one repetition
of the nine–tone multitone waveform segment.
a. Press
Mode > Dual ARB > Waveform Sequences > Build New Waveform Sequence > Insert Waveform.
b. Highlight the a waveform segment (for example, TTONE) and press
Insert Selected Waveform.
c. Highlight a second waveform segment (for example, MTONE) and press
Insert Selected Waveform.
d. Press
Done Inserting
2. Optional: Enable markers as desired for the segments in the new sequence: see page 106.
3. Name and store the waveform sequence to the Catalog of Seq Files in the memory catalog:
a. Press
Name and Store.
b. Enter a file name (for example, TTONE+MTONE).
c. Press
Enter.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8257D PSG and is the answer not in the manual?

Agilent Technologies E8257D PSG Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8257D PSG
CategoryInverter
LanguageEnglish

Related product manuals