1. The result needs to have two peaks (more than two or less than two will
also failed) and the S/N must be 60 times of noise.
2. Peak time must be within the specified range, cannot be shift too far.
If the calibration failed, user can do it again, and it will extend 10s of the HV
purge (10n, n= calibration number) and 30s of separation in next
calibration.
The cartridge will be locked after 10 times failed.
*Note: make sure user place above 20µl of 20bp& 1kb in MA1 (M) position properly.




01
RFID module is
out of order
RFID Reader
Re-latch
02
Can’t find the Tag
RFID Reader or
RFID tag
Re-latch
Use the other cartridge to confirm
05
Incorrect format
of RFID Tag
RFID tag
Re-latch
06
Error from RFID
decoding
RFID tag
Re-latch



Using the purge station to see whether the gel drop comes out. If not, soak
the cartridge tip in the beaker with hot water for 10-15mins and then purge
again.
It will be even better to soak and purge at the same time

Table of Contents

Other manuals for BiOptic Qsep100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BiOptic Qsep100 and is the answer not in the manual?

BiOptic Qsep100 Specifications

General IconGeneral
BrandBiOptic
ModelQsep100
CategoryMedical Equipment
LanguageEnglish

Related product manuals