Qair
box

Check the air leaking of the system airway. You can also use the bubbling water
to check it.
We have used three kinds of filters. The right one is the strongest, which can be
used longer than the other two. And the middle white one is stronger than the
left blue one.
the cartridge cap locks too tight or too loose may also cause air leaks.
1. Too tight: The distance between the cartridge cap and the purge piston will
be enlarged.
2. Too loose: The distance between the cartridge cap and gel reservoir will be
enlarged.

Please send all the files mentioned in the warning window back to the BiOptic.